AMETEK Programmable Power MX / RS Series SCPI Programming Manual
186 MX / RS Series
[:SOURce]:IHARmonics:REFerence
This command will program the inter harmonic reference voltage. The coupling must be set
to on to accept any program value. The reference will take a value of 0 to 250V.
Command Syntax [:SOURce]:IHARmonics:REFerence <NRf>
Parameters <reference>
Examples IHAR:REF 230
Query Syntax IHAR:REF?
Returned Parameters <NR1>
[:SOURce]:IHARmonics:VOLTage
This command will program the interharmonic voltage as a percentage of the reference
voltage. It will accept a value from 0 to 20%.
Command Syntax [:SOURce]:IHARmonics:VOLTage <NRf>
Parameters <voltage>
Examples IHAR:VOLT 10
Query Syntax IHAR:VOLT?
Returned Parameters <NR1>