California Instruments 801RP Series Spécifications Page 117

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 126
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 116
To measure the RMS current, use:
MEAS:CURR?
To measure the peak current, use:
MEAS:CURR:AMPL:RES /* resets track and hold for peak curr. meas.
/* It may be need to turn the output on or program
/* as specific voltage at this time before taking a
/* peak current reading.
MEAS:CURR:AMPL:MAX? /* retrieves peak current reading.
To measure the current crest factor, use:
MEAS:CURR:CRES?
Power Measurements
To measure the true power, use:
MEAS:POW:REAL?
To measure the apparent power, use:
MEAS:POW:APP?
To measure the power factor, use:
MEAS:POW:PFAC?